Figure 3 and 4 indicate good separation
performances in the soil extract and
water sample. Obtained results from
CE-C4D are in good agreement with
those from the confirmation method (FAAS for K+ analysis and UV-Vis for
NH4+ analysis) (Table 4). This suggests
that the conditions found are applicable
for detection of NH4+ and K+ in
environmental samples. Levels of NH4-
N and K+ determined in the real soil
sample were of 29.6 and 1055 mg/kg
dried soil, respectively. Those highly
exceeded the normal level of NH4-N
(10 mg/kg) and the excessive level of K
(> 800mg/kg) defined by D.A. Horneck
et al. [11]. The concentration of NH4-N
in water sample was more than 5 times
higher than the National technical
regulation on surface water quality for
protection of aquatic lives [12]. It
should be noted that sufficiently high
levels of NH4+ and K+ in agricultural
run-off and soil would cause nutrient
imbalance and serious environmental
problem
9 trang |
Chia sẻ: honghp95 | Lượt xem: 518 | Lượt tải: 0
Bạn đang xem nội dung tài liệu Determination of ammonium and potassium extracted from soil samples by capillary electrophoresis with capacitively coupled contactless conductivity detection (ce-C4d) - Van Anh Nguyen, để tải tài liệu về máy bạn click vào nút DOWNLOAD ở trên
135
Tạp chí phân tích Hóa, Lý và Sinh học - Tập 21, Số 2/2016
DETERMINATION OF AMMONIUM AND POTASSIUM EXTRACTED
FROM SOIL SAMPLES BY CAPILLARY ELECTROPHORESIS WITH
CAPACITIVELY COUPLED CONTACTLESS CONDUCTIVITY
DETECTION (CE-C4D)
Đến tòa soạn 23 - 3 - 2016
Van Anh Nguyen
Department of Environmental Sciences and Technologies, Hanoi Metropolitan
University
Thi Trang Do, Van Lau Ha, Hung Viet Duong, Tien Duc Pham,
Thi Anh Huong Nguyen*
Faculty of Chemistry - Hanoi University of Science - Vietnam National University,
Hanoi,
Thu Hien Le
Faculty of Transport Economics, University of Transport Technology
TÓM TẮT
XÁC ĐỊNH AMONI VÀ KALI TRONG DỊCH CHIẾT CỦA MẪU ĐẤT
BẰNG PHƯƠNG PHÁP ĐIỆN DI MAO QUẢN SỬ DỤNG DETECTOR ĐỘ
DẪN KHÔNG TIẾP XÚC (CE-C4D)
Bài báo nghiên cứu quy trình phân tích hai ion NH4+ và K+ trong dịch chiết mẫu đất
nông nghiệp bằng kĩ thuật điện di mao quản sử dụng detector độ dẫn không tiếp xúc.
Các điều kiện phân tích đã được tối ưu bao gồm: Dung dịch đệm điện di sử dụng hỗn
hợp đệm histidine/axit acetic (His/Ace) (pH = 4), 2mM 18-crown-6; mao quản Silica:
đường kính ngoài (O.D.) 365 μm, đường kính trong (I.D.) 50 μm, chiều dài tổng Lt=
50cm và chiều dài hiệu dụng Leff = 40cm; điện thế tách: 20kV; bơm mẫu theo phương
pháp thủy động lực theo kiểu xiphong ở độ cao 15cm, trong thời gian 20s. Quy trình
được áp dụng để phân tích dịch chiết 01 mẫu đất và 01 mẫu nước lấy tại cùng một
khu canh tác nông nghiệp. Các kết quả phân tích được kiểm chứng bằng các phương
136
pháp quang phổ hấp thụ nguyên tử (kiểm chứng kết quả phân tích K+) và phương
pháp quang phổ hấp thu phân tử (kiểm chứng kết quả phân tích NH4+). Sai số tương
đối giữa phương pháp phân tích và các phương pháp đo kiểm chứng đều dưới 15%.
Kết quả cho thấy hàm lượng NH4-N và K+ trong mẫu đất lần lượt là 29,6 và 1055
mg/kg đất khô. Hàm lượng NH4-N và K+ trong mẫu nước lần lượt là 2,68 và 33,7
mg/L. Hàm lượng NH4-N trong mẫu nước ruộng vượt hơn 5 lần quy chuẩn kĩ thuật
Quốc gia về chất lượng nước mặt bảo vệ đời sống thủy sinh.
Key words: ammonium, potassium, contactless conductivity detector, agricultural soil
1. INTRODUCTION
Nitrogen and potassium are essential
nutrients for plant growth, survival,
development and reproduction.
However, there are not always enough
of these nutrients in the soil for a plant
to grow healthily so that many farmers
and gardeners use fertilizers as the
nutrients for the soil. Despite of the fact
that fertilizers are beneficial to
agriculture, excessive or improper
fertilizer applying would induce to
environmental and human health risks.
The excess of nutrients in environment
would cause soil pollution, surface
water eutrophication, groundwater
contamination etc. Elevated levels of
nutrients accumulated in crop products
could make unexpected symptoms and
diseases to consumers [1]. For better
agricultural land use, it is necessary to
quantify the nutrient residues that are
available to be released to environment.
The extractions of the exchangeable
forms of K+ and NH4+ can be performed
by using salt solutions. Because NH4+
and K+ exhibit nearly identical
characteristics, common extracting
reagents for exchangeable potassium
are ammonium salts. On the other hand,
ammonium adsorbed on the exchange
complex is normally removed by using
potassium salts such as KCl or K2SO4
[2]. Thus, measurements of soil extracts
for K+ and NH4+ would be strongly
interfered by the matrix effects.
Capillary Electrophoresis (CE) with
Capacitively Coupled Contactless
Conductivity Detection (C4D) is a
simple and inexpensive method that can
be applicable to simultaneously
determine NH4+ and K+ in aqueous
samples. However, the separations of
two cations solely based on differences
in their electrophoretic mobility seem to
be a challenge.
Numerous attempts were investigated to
separate NH4+ and K+ in aqueous
solutions. One of the most successful
approach was the employment of
complex-forming reactions. Among
comlexing agents, 18-crown-6 ether
137
was widely used to separate the NH4+
and alkaline metal cations as it forms a
stable complex with alkaline metal
cations [3, 4, 5, 6, 7]. Adam J. Gaudry
et al. separated an aqueous mixture
comprising three inorganic cations
(NH4+, Na+ and Li+) in 50 mM acetic
acid/10 mM L-histidine /2.5 mM18-
crown-6 ether electrolyte at pH 4.2. The
limit of detections (LODs) of NH4+,
Na+, and Li+ were 1.54, 2.26, and
3.06mg/L, respectively [3]. Another
separation of alkaline metal cations (K+,
Na+, Li+) in aqueous solutions was
obtained by using the background
electrolyte (BGE) of 12 mM histidine
adjusted to pH 4 with acetic acid and 2
mM 18-crown-6. The LOD for K+ was
of 1.5 μM [8]. Solutions containing
NH4+, K+ and other cations were
determined by Thanh Duc Mai et al.
[5]. Background electrolyte solutions
were His 12 mM (adjusted to pH 4 with
acetic acid) in the presence of 2 mM of
18-crown-6; capillary was of 50 μm
I.D., 36 cm effective length, and 50 cm
total length, while separation voltage
was 15 kV. The LODs were in the
lower micromolar range and varied
depending on species. Baseline
separation between NH4+ and K+ was
achieved at the concentration of less
than100 μM for each cation [5].
Nevertheless, to our best knowledge, a
simultaneous separation of NH4+ and K+
aqueous solutions extracted from soils
by CE-C4D has not been reported.
In the present study, operation
parameters were set up in our laboratory
according to previously published
papers for measuring NH4+ and K+ in
aqueous solutions. The optimizing
amount of 18-crown-6 was
systematically investigated. Appropriate
extracting solutions used to remove
exchangeable faction of each ion from
soil were also examined. The
preliminary conditions were applied to
detect NH4+ and K+ in a water sample
and the extracts of a soil sample
collected from agricultural land in
Vietnam.
2. MATERIALS AND METHODS
2.1. Materials
All the chemicals such as NH4Cl,
K2SO4, His, C6H9N302, HCl, NaOH
etc were of analytical grade and
purchased from Fluka or Merck.
Ultrapure water system (Labconco,
USA) with resistivity 18.2 MΩ was
used to produce ultrapure water in all
solutions and measurements. Stock
solutions of ammonium and potassium
ions were daily prepared. The
background electrolyte (BGE) solutions
were prepared with the buffer of
histidine and acetic acid (His/Ace)
(pH=4) in the presence of 18-Crown-6.
The pH of solutions was controlled by
using an HI 2215 Hanna Instruments
138
pH meter (Woonsocket, RI, USA).
Fused silica capillaries of 50 μm I.D.
and 365 μm O.D. with total length (Lt)
of 50 cm and effective length (Leff) of
40 cm (purchased from Polymicro,
Phoenix, AZ, USA) were used for
separations. The capillaries were
preconditioned with 1 M NaOH for
10minutes and then, deionized water for
10minutes prior to flushing with buffer
solutions. Experiments were performed
using the portable semi-automated CE
system supported by 3Sanalysis JSC.
( (Figure 1).
More detailed information about the
instrument was given elsewhere in T. A.
H. Nguyen et. al. [9].
Figure 1. Portable semi-automated CE-C4D instrument
2.2. Methods
According to former references
reviewed earlier, the BGE with the
buffer adjusted to pH 4 was commonly
used for the separations of the two
cations using CE -C4D. In the present
study, the buffer solution of His/Ace
(pH=4) was selected. The high voltage
of 20kV was applied for separation.
Hydrodynamic injection of samples was
carried out by setting the high-voltage
end of the capillary at 15cm height. The
injection time was 20 seconds. Table 1
summarized the operating parameters
used in this research.
139
Table 1. Separation conditions of CE-C4D system for determine NH4+ and K+.
BGE composition His/Ace buffer adjusted to pH 4
18-Crown-6
High voltage 20kV
Detection C4D
Capillary 365 μm O.D., 50 μm I.D.; Lt= 50cm; Leff=
40cm; Preconditioned
Hydrodynamic injection 20s at 15cm height
The BGE solutions with differences
amount of 18-crown-6 were prepared in
order to get optimum BGE composition
for the separation of two cations. The
concentrations of 18-crown-6 varied
from 1 to 3mM. The concentrations of
both NH4+ and K+ were 0,1mM.
Appropriate soil extraction procedure
was investigated. Simulated soil
samples contaminated by 20mM of K+
and 2.2mM of NH4+ were prepared by
adding certain amount of K+ and NH4+
solutions to uncontaminated soils. The
samples were left for 1 day to reach to
equilibrium, and then dried in
desiccator before any further
preparation. For K+ extraction, the
CH3COONH4 solutions of 0.01M and
0.1M were used. The ion NH4+ was
extracted with K2SO4 solutions of 0.001
and 0.1 M. The amount of 1.000 g dried
soil and precisely 20ml of extracting
agent were put into 50ml- falcon tubes
and shaken for 1 hour using a shaker
(Cole Parmer, 51704 Series). Finally,
the mixture was centrifuged and filtered
through 0.45μm membrane. The
filtrates were used for the K+ and NH4+
measurements using CE-C4D. The
levels of K+ and NH4+ were compared
with Flame - Atomic Absorption
Spectroscopy (F-AAS) and Ultraviolet–
Visible Spectroscopy ((UV-Vis),
respectively.
One soil and one water samples were
collected in agricultural regions Thuong
Tin, Hanoi, Vietnam. The soil sample
was immediately stored at 4oC and then
dried in desiccator. The water sample
was kept at 4oC and filtered through
0.45μm membrane before analysis.
3. RESULTS AND DISCUSSION
It was known that potassium forms a
stable complex with 18-crown-6 [10].
Consequently, the presence of 18-
crown-6 in the BGE increases the
migration time of K+ while the
migration time of NH4+ stays
unchanged. As clearly shown in Figure
2, the best separation was performed at
the concentration of 2mM 18-crown-6.
Lower concentrations than that of 2 mM
140
18-crown-6 would not be enough for
reasonable separation of the two cations
while higher concentrations seem to be
not necessary. Thus, for further
experiments the BGE solutions contain
Tris/Ace buffer (pH=4); 2 mM 18-
Crown-6.
15014013012011010090
10 mV
Migration time (s)
1
2
3
4
5
6
Figure 2. Separation of NH4+ and K+ with BGE solutions containing different levels of 18-
crown-6; Concentration of NH4+ and K+: 10-4M; Concentration of 18-crown-6: (1): Non
18-crown-6 contained; (2) 1.0 mM; (3) 1.5 mM; (4) 2.0 mM; (5): 2.5 mM. (6): 3 mM
The calibration curves (six points) were
extrapolated using standard addition
method in order to avoid the matrix
interference of the sample (Table 2).
Linear range extended to the
concentration of 200 μM for each
cations with a correlation coefficient
(r2) of at least 0.99. The LOD and LOQ
values were calculated from peak areas
corresponding to 3 and 10 times the
baseline noise (S/N= 3 and S/N= 10),
respectively. The LOD values (1.0 µM
for NH4+ and 3.0 µM for K+) are
comparable to previously reported
results [5, 8].
Table 2. Calibration curve extrapolated by standard addition method
Analyte Rangea (μM) r2 LODb (μM) LOQc (μM)
NH4+ 50.0-200 0.9988 1.0 5.0
K+ 50.0-200 0.9909 3.0 10.0
a Six points
b S/N: 3
c S/N: 10
As can be seen in Table 3, the best
extraction efficiency was achieved
when using 0.1M CH3COONH4 and
0.001M K2SO4 for extraction K+ and
141
NH4+, respectively. Higher
concentrations of extractants are
unfavorable due to the strong matrix
effect to the separation of the two
cations by CE, while lower ones are not
enough to completely extract the
exchangeable solutes.
Table 3. Extraction of the simulated soil sample for K+ and NH4+
Extractant Extraction efficiency (%)
Extraction for K+ CH3COONH4 0.01M 25.0
CH3COONH4 0.1M 99.2
Extraction for NH4+ K2SO4 0.001M 109.0
K2SO4 0.1M -a
aunable to detect
Figure 3 and 4 indicate good separation
performances in the soil extract and
water sample. Obtained results from
CE-C4D are in good agreement with
those from the confirmation method (F-
AAS for K+ analysis and UV-Vis for
NH4+ analysis) (Table 4). This suggests
that the conditions found are applicable
for detection of NH4+ and K+ in
environmental samples. Levels of NH4-
N and K+ determined in the real soil
sample were of 29.6 and 1055 mg/kg
dried soil, respectively. Those highly
exceeded the normal level of NH4-N
(10 mg/kg) and the excessive level of K
(> 800mg/kg) defined by D.A. Horneck
et al. [11]. The concentration of NH4-N
in water sample was more than 5 times
higher than the National technical
regulation on surface water quality for
protection of aquatic lives [12]. It
should be noted that sufficiently high
levels of NH4+ and K+ in agricultural
run-off and soil would cause nutrient
imbalance and serious environmental
problem.
150140130120110100
10 mV
Migration time (s)
NH4
+
K
+
1
2
Figure 3. Separation of NH4+ and K+ in the extracts of soil sample:
(1) Soil extract for NH4+; (2) Soil extract for K+, dilution factor (DF) was 50 times
142
150140130120110100
10 mV
Migration time (s)
NH4
+
K
+
Figure 4. Simultaneous determination of NH4+ and K+
in water sample (DF was 6 times)
Table 4. Concentrations of NH4+ and K+ in soil and water samples
Type of sample Analyte Concentration* Confirmation
(% difference)
Soil sample K+ 1055 5,7a
NH4-N 29.6 10.8b
Water sample K+ 33,7 12,9a
NH4-N 2.68 0.50b
* mg/kg for soil sample; mg/L for water sample
a confirmation by F-AAS method
b confirmation by UV-Vis method
4. CONCLUSIONS
The conditions for simultaneous
determination of NH4+ and K+ by CE-
C4D were optimized as follows: BGE:
His/Ace buffer, 2 mM 18-crown-6;
silica capillary: 365 μm O.D., 50 μm
I.D.; Lt = 50cm; Leff = 40cm; voltage:
20V; hydrodynamic injection: 20s at
15cm height. The procedure are
applicable to examine NH4+ and K+ in
both soil and water samples. Excessive
levels of NH4+ and K+ were detected in
soil. The concentration of NH4+ in water
sample was significantly higher than the
National technical regulation on surface
water quality for protection of aquatic
lives.
ACKNOWLEDGEMENT
This research was partly funded by
Hanoi Metropolitan University under
the project number C.2015-19.
REFERENCE
[1] Stanley E. Manahan, (2009),
Environmental chemistry, CRC Press.
[2] Marc Pansu, Jacques Gautheyrou,
(2006), Handbook of Soil Analysis:
Mineralogical, Organic and Inorganic
Methods, Springer.
[3] Adam J. Gaudry, Michael C.
Breadmore, and Rosanne M. Guijt
(2013), In-plane alloy electrodes for
143
capacitively coupled contactless
conductivity detection in poly
(methylmethacrylate) electrophoretic
chips, Electrophoresis, 34, 2980–2987.
[4] Marko Stojkovic, Boris Schlensky,
Peter C. Hauser, (2013), Referenced
Capacitively Coupled Conductivity
Detector for Capillary Electrophoresis,
Electroanalysis, 25, No. 12, 2645 –
2650
[5] Thanh Duc Mai, Thi Thanh Thuy
Pham, Hung Viet Pham, Jorge Saiz,
Carmen García Ruiz and Peter C.
Hauser, (2013), Portable capillary
electrophoresis instrument with
automated injector and contactless
conductivity detection, Anal. Chem.,
85, 2333−2339
[6] Thanh Duc Mai, Stefan Schmida,
Beat Müller, Peter C. Hauser (2010),
Capillary electrophoresis with
contactless conductivity detection
coupled to asequential injection analysis
manifold for extended automated
monitoring applications, Analytica
Chimica Acta, 665, 1–6
[7] Pavel Kubáň and Peter C. Hauser
(2009), Ten years of axial capacitively
coupled contactless conductivity
detection for CZE - a review.
Electrophoresis, 30, pp.176-188.
[8] Adam J. Gaudrya, Rosanne M.
Guijtb, MirekMackaa, Joseph P.
Hutchinsona, CameronJohnsa, Emily
F. Hildera, Greg W. Dicinoski, Pavel
N. Nesterenkoa, Paul R. Haddada,
Michael C. Breadmorea, (2013), On-
line simultaneous and rapid separation
of anions and cations from a single
sample using dual-capillary sequential
injection-capillary electrophoresis,
Analytica Chimica Acta, 781, 80– 87.
[9] Thi Anh Huong Nguyen, Thi
Ngoc Mai Pham, Thi Tuoi Doan, Thi
Thao Ta, Jorge Sáizc, Thi Quynh Hoa
Nguyen, Peter C. Hauser, Thanh Duc
Mai, (2014), Simple semi-automated
portable capillary electrophoresis
instrument with contactless
conductivity detection for the
determination of β-agonists in
pharmaceutical and pig-feed samples,
Journal of Chromatography A, 1360,
305–311.
[10] Harald Heiland, John A. Ringseth
and Thorvald S. Brun (1979), Cation-
Crown Ether Complex Formation in
Water. II. Alkali and Alkaline Earth
Cations and 12-Crown-4, 15-Crown-5,
and 18-Crown-6, Journal of Solution
Chemistry, Vol. 8, No. 11.
[11] D. A. Horneck, D.M. Sullivan,
J.S. Owen, and J.M. Hart (2011), Soil
Test Interpretation Guide, EC 1478.
Corvallis.
[12] Ministry of Science, Technology
and Environment (2011) QCVN
38:2011/BTNMT, National technical
regulation on surface water quality for
protection of aquatic lives.
Các file đính kèm theo tài liệu này:
- 26480_89041_1_pb_3381_2096853.pdf