CONCLUSIONS
An international code of practice for the
dosimetry of small static fields used in external
beam radiotherapy (TRS483 CoP) was
successfully applied to recalculate relative
output factors for cone system with correction.
Relative output factors for jaw collimation
system were extensively obtained for field size
less than 3 × 3 cm2 for Eclipse v.13.6 for 6X
and 6XFFF beams using TRS483 CoP.
Relative output factors were also measured for
MLC collimation system to be compared with
that of the jaw collimation system. The
discrepancy of output factor between jawshaped fields and MLC-shaped fields suggests
that jaw-based beam data itself may not
suitable for MLC-based treatment planning.
Additional measurement of small beam
percentage depth dose and profiles as well as
specific modelling of photon beam for MLC
system may be required.
8 trang |
Chia sẻ: hachi492 | Lượt xem: 15 | Lượt tải: 0
Bạn đang xem nội dung tài liệu Relative output factors of different collimation systems in truebeam STx medical linear accelerator, để tải tài liệu về máy bạn click vào nút DOWNLOAD ở trên
Nuclear Science and Technology, Vol.9, No. 4 (2019), pp. 48-55
©2019 Vietnam Atomic Energy Society and Vietnam Atomic Energy Institute
Relative output factors of different collimation systems in
truebeam STx medical linear accelerator
Do Duc Chi
1
, Tran Ngoc Toan
2
, Robin Hill
3,4
, Nguyen Do Kien
1
1
108 Military Central Hospital, Hanoi, Vietnam
2
Vietnam Atomic Energy Institute, Hanoi, Vietnam
3Chris O’Brien Lifehouse, NSW, Australia
4
School of Physics, The University of Sydney, NSW Australia
Email: chidd108@gmail.com
(Received 05 September 2019, accepted 25 September 2019)
Abstract: The IAEA TRS483 and TRS398 Code of Practices (CoP) were used to calculate
relative output factors for small photon beams of 6X, 6XFFF energies shaped by High
Definition Multileaf Collimator (HDMLC), jaws and cones mounted on TrueBeam STx
medical linear accelerator (Varian Medical Systems), respectively. A comparison between
these results were made. The results show a large discrepancy in relative output factor
curves found among different collimation systems of the same equivalent field sizes and
between the CoPs. Therefore, the specific beam modelling in treatment planning system
for each type of the collimation system to be used for small fields maybe required for
better computational accuracy.
Keywords: TRS483 code of practice, small field dosimetry, relative output factors.
I. INTRODUCTION
Modern radiotherapy techniques such
as Intensity-Modulated Radiation Therapy
(IMRT), Volumetric Modulated Arc
Therapy (VMAT), Stereotactic
Radiosurgery (SRS) and Stereotactic
Radiation Therapy (SRT) make use of small
photon beams in order to deliver complex
radiation treatments. However, there are
still many physical and technical aspects
which need to be considered in order to
commission small photon beams safely and
efficiently in clinical practice such as:
changing in photon fluence spectrum
making beam quality changing by field size,
lateral disequilibrium of charged particles
may leading to wrong estimation of
absorbed dose as well as detector size
compared to field size [1–4].
Practical issues encountered are:
photon beam data for treatment planning
system (TPS) are usually collected for jaw-
shaped beams while we use these data for
computation of Multileaf Collimator
(MLC) - shaped beams. Furthermore,
HDMLC-shaped beams are constituted
from very tiny beamlets, much smaller than
smallest collected beam data of field size
of 3 × 3 cm
2
(at isocenter), which may
affect the computation accuracy of TPS,
especially for small tumors. In Eclipse
v.13.6 (Varian Medical Systems), warning
message “inaccuracy” was often seen when
making treatment plans for tumors less
than 3 cm diameter. Radiation oncologists
tend to use HDMLC for small tumor
radiosurgery because of its small thickness
(2.5 mm at isocenter) and convenience.
DO DUC CHI et al.
49
Fig. 1. Occlusion of photon source in the case of narrow collimation. Left: the full, extended source can be
“viewed” by an observer on the central axis. Right: only partial view of the source is possible by an observer
on the central axis [13].
Conical collimators are dedicated for
radiosurgery of small tumors. With cone-
shaped beams, field size diameters are of
17.5 mm down to 4 mm cone, but they are
previously measured using TRS 398 CoP
published by the IAEA [14]. It has been
shown that the beam quality of photon
beam changes significantly due to these
very small field collimations [2], [5–9]. In
this study, we made a comparison of
relative output factors of different
collimation systems (jaws, HDMLC and
cones) for further estimation of the
computational accuracy of TrueBeamSTx
TPS using newly published TRS 483 CoP
by the IAEA [2].
Fig. 2. TrueBeam STx treatment head with collimation systems:
a) Jaws (highest), HDMLC (midlle) [10] and b) Cone (lowest)
RELATIVE OUTPUT FACTORS OF DIFFERENT COLLIMATION SYSTEMS IN TRUEBEAM
50
II. MATERIAL AND METHOD
TrueBeamSTx medical linear accelerator
with integrated HDMLC with 20 central leaf
pairs of 2.5 mm thickness and 40 peripheral
leaf pairs of 5.0 mm thickness at isocenter.
Beam shaping using HDMLC (and also jaws)
were of field sizes 0.5 × 0.5, 1 × 1, 2 × 2, 3 × 3,
4 × 4, 5 × 5, 7 × 7, 10 × 10 cm
2
. MLC-shaped
fields were created when the jaws were
“optimized” and at “recommended positions”
by software. Inversely, jaw-shaped fields were
created when MLC are fully retracted. Beam
shaping using the cones were with diameter of
4.0, 5.0, 7.5, 10.0, 12.5, 15.0 and 17.5 mm.
Photon beam energies of 6X (with Flatterning
Filter), 6XFFF (Flatterning Filter-Free), 10X,
10XFFF were used for measurements. The
linac was calibrated for all photon energies at
10 × 10 cm
2
jaw-shaped field to be used for all
other collimation systems.
The dose measurements were
performed in Blue Phantom 2 (IBA) using a
Razor chamber (IBA) and Razor diode (IBA)
under Source-to-Axis Distance setup (100 cm
SAD, 5 cm depth). The TRS398 and TRS483
CoP are both applied to determine relative
output factors. Relative output factor curves
were compared for 3 different collimation
systems and in both CoPs. All data were
normalized to 10× 10 cm
2
field size. The
equivalent square of the cone defined fields
were calculated using formula [2] :
√ √ (1)
Razor diode (unshielded, p-type silicon
diode chip, active detector diameter of 0.6 mm)
with high spatial resolution and high sensitivity
is superior to Razor chamber (total active
length of 3.6 mm) in relative dosimetry of
small photon beams. However, Razor diode
has an over-response in large fields because of
the significant amount of phantom scatter
component of low energy photons. The
consequence is an underestimation of field
output factors when they are normalized to a
large field size (e.g. the conventional 10 cm ×
10 cm
2
reference field) [2].
According to TRS398 CoP, the output
factor may be determined as the ratio of
corrected dosimeter readings measured under
a given set of non-reference conditions to
that measured under reference conditions.
However, in TRS483 CoP, the field output
factor,
, relative to is defined
by the following equation :
(2)
Where
and
are the
readings of the detector (corrected for
influence quantities) in the clinical field (fclin)
and the machine specific reference field (fmsr),
respectively.
is a beam quality
correction factor which changed by field size.
The intermediate field ( ) method
was used by applying formula (2) for field
sizes bigger than measured by Razor
ionization chamber, and field sizes smaller
than measured by Razor unshielded diode
for using formula (2):
[
]
[
]
(3)
where “det” refers to the small field
detector (Razor diode) and “IC” to the
ionization chamber (Razor chamber). The
output correction factor [
]
is
a
DO DUC CHI et al.
51
obtained from the tabulated output correction
factors with respect to the machine specific
reference field as below:
[
]
[
]
[
]
(4)
III. RESULTS and DISCUSSION
A. Output factors of collimation systems
using TRS398 CoP
Using the conventional formula from
TRS398 CoP, we got the result as Table I.
Table I. Output factors of collimation systems using TRS398 CoP and Razor chamber.
Cone (mm) 4 5 7.5 10 12.5 15 17.5 100 (square)
6X 0.420 0.523 0.662 0.746 0.797 0.834 0.859 1
6XFFF 0.473 0.574 0.702 0.772 0.816 0.846 0.866 1
MLC FS (mm) 0.5 1 2 3 4 5 7 10
6X 0.529 0.754 0.861 0.895 0.921 0.940 0.968 1
6XFFF 0.560 0.782 0.876 0.909 0.932 0.948 0.975 1
Jaw FS (mm) 0.5 1 2 3 4 5 7 10
6X 0.345 0.704 0.850 0.888 0.914 0.936 0.967 1
6XFFF 0.377 0.736 0.867 0.906 0.929 0.947 0.974 1
B. Output factor of collimation systems
using TRS483 CoP:
Intermediate field ( ) of 4 × 4
cm
2
was selected for calculation of jaw-
shaped fields and MLC-shaped fields. For
cone, intermediate field was 17.5 mm
conical field because we need to
normalize these data to that of 10 × 10
cm
2
field size. The results were obtained
as Table II.
Table II. Output factor of collimation systems using TRS483 CoP (Razor chamber and Razor diode)
Cone (mm) 4 5 7.5 10 12.5 15 17.5 100 (square)
Square Equi. Field
size (cm) 0.708 0.885 1.327 1.77 2.212 2.655 3.097 10
6X 0.522 0.599 0.713 0.773 0.814 0.841 0.864 1
6XFFF 0.578 0.648 0.745 0.797 0.830 0.856 0.871 1
MLC Field size(mm) 0.5 1 2 3 4 5 7 10
6X 0.608 0.774 0.871 0.905 0.927 0.945 0.971 1
6XFFF 0.643 0.792 0.881 0.918 0.938 0.954 0.978 1
Jaw Field size(mm) 0.5 1 2 3 4 5 7 10
6X 0.619 0.756 0.84 0.876 0.901 0.924 0.961 1
6XFFF 0.652 0.771 0.846 0.884 0.907 0.928 0.963 1
C. Comparison of results between TRS483
and TRS398 CoP:
Based on these results, the difference
between ROF curves is significant between the
two different methods (CoPs) for MLC-shaped
field size and for jaw-shaped field size less than
3 × 3 cm for both the 6X and 6XFFF beams.
The smallest difference was observed
with MLC-shaped fields while the biggest
difference was observed with cone-shaped
RELATIVE OUTPUT FACTORS OF DIFFERENT COLLIMATION SYSTEMS IN TRUEBEAM
52
fields as seen in Fig.3 and Fig.4. At 0.5 × 0.5
cm
2
squared field and 4 mm conical field, the
output factor difference of 6X and 6XFFF
beams were -44.2%/-42.2%, -13.0%/-13.0%, -
19.6%/-18.2% for jaw-shaped, MLC-shaped
and cone-shaped fields, respectively. TRS398
CoP gave underestimation of a relative output
factor in comparison with TRS483 CoP. The
large difference was always seen at field sizes
smaller than 4 × 4 cm
2
.
Fig. 3. Difference of TRS398 and TRS483 CoP in relative output factor of MLC and Jaws collimations.
Fig. 4. Difference of TRS398 and TRS483 CoP in relative output factor of cone collimations.
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
0 2 4 6 8 10
R
O
F
Jaw Field Size (cm x cm)
6X Jaw Output Factors
TRS483
TRS398
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
0 2 4 6 8 10
R
O
F
Jaw Field Size (cm x cm)
6XFFF Jaw Output Factors
TRS483
TRS398
0.5
0.6
0.7
0.8
0.9
1
0 2 4 6 8 10
R
O
F
MLC Field Size (cm x cm)
6X MLC Output Factors
Razor Chamber -
TRS398
0.5
0.6
0.7
0.8
0.9
1
0 2 4 6 8 10
R
O
F
MLC Field Size (cm x cm)
6XFFF MLC Output Factors
Razor Chamber -
TRS398
TRS483
0.4
0.5
0.6
0.7
0.8
0.9
1
0 10 20 30 40 50 60 70 80 90 100
R
O
F
Coned Field Size (mm)
6X Cone Output Factors
TRS483
TRS398
0.4
0.5
0.6
0.7
0.8
0.9
1
0 10 20 30 40 50 60 70 80 90 100
R
O
F
Coned Field Size (mm)
6XFFF Cone Output Factors
TRS483
TRS398
DO DUC CHI et al.
53
Noticingly, the Razor chamber’s
reading differences between 10 × 10 cm
2
MLC-shaped field and 10 × 10 cm
2
jaw-
shaped field were just 0.61% and 0.25% for
6X and 6XFFF, respectively. Therefore, these
relative output factors could be used for direct
comparison between jaw-shaped field and
MLC-shaped field of “the same” nominal
field size.
D. Comparison of results between 6X,
6XFFF (TRS483 CoP):
For the same collimation system, output
factor comparisons were also made for 6X and
6XFFF beams after applying TRS483 CoP
(Fig. 5). The biggest differences in output
factor were seen at 0.5 × 0.5 cm
2
jaw-shaped
field, 0.5 × 0.5 cm
2
MLC-shaped field and
4mm cone-shaped field with values of 5.3%,
5.8% and 10.5%, respectively.
Fig. 5. Difference in output factor of 6X and 6XFFF beams in each collimation system.
E. Comparison of Output Factor curves
between different collimation systems
(TRS483 CoP):
The relative output factor comparisons
were made between MLC, jaws and Cone
systems for both 6X and 6XFFF beams.
Conical collimators are independent
from MLC and Jaws systems. The conical
collimation system has smallest relative
output factor in comparison with that of MLC
and Jaws systems for both 6X and 6XFFF
beams as Fig. 6.
For field sizes bigger than 1 × 1 cm
2
, jaw
system has lower relative output factor than
MLC’s but it is inverse for field size less than 1
× 1 cm
2
.
In a multi-centre analytical study of
small field output factor calculations in
radiotherapy reported by Krzysztof Chełmiński
and Wojciech Bulski, for 2 × 2 cm
2
MLC-
shaped fields of Varian linacs, the differences
between the treatment planning system output
factors (based on collected beam data) often
exceeded 5% and were below 10% [11]. In
our study, these differences were -1.1% (6X)
and -0.5% (6XFFF) for MLC-shaped fields,
1.3% (6X) and 2.6% (6XFFF) for jaw-shaped
fields, -7.0% (6X) and -5.7% (6XFFF) for
cone-shaped fields. The smaller differences
0.5
0.6
0.7
0.8
0.9
1
0 2 4 6 8 10
R
O
F
Jaw Field Size (cm)
TRS483 Jaws ROF
6X
6XFFF
0.5
0.6
0.7
0.8
0.9
1
0 2 4 6 8 10
R
O
F
MLC Field Size (cm)
TRS483 MLC ROF
6X
6XFFF
0.5
0.6
0.7
0.8
0.9
1
2.5 5 7.5 10 12.5 15 17.5
R
O
F
Coned Field Size (mm)
TRS483 Cone ROF
6X
6XFFF
RELATIVE OUTPUT FACTORS OF DIFFERENT COLLIMATION SYSTEMS IN TRUEBEAM
54
observed in our study for MLC-shaped field
may came from our small field detector, the
Razor chamber.
A multinational audit of small field
output factors calculated by treatment
planning systems used in radiotherapy, the
ROF for small fields calculated by TPSs
were generally larger than measured
reference data. On a national level, 30% and
31% of the calculated ROF of the 2 × 2 cm
2
field exceeded the action limit of 3% for
nominal beam energies of 6 MV and for
nominal beam energies higher than 6 MV,
respectively [12].
The discrepancy above may come
from accuracy of treatment planning
algorithms on measured output factors,
especially for small fields.
Fig. 6. Difference in output factor of difference collimation system for 6X and 6XFFF beams.
CONCLUSIONS
An international code of practice for the
dosimetry of small static fields used in external
beam radiotherapy (TRS483 CoP) was
successfully applied to recalculate relative
output factors for cone system with correction.
Relative output factors for jaw collimation
system were extensively obtained for field size
less than 3 × 3 cm
2
for Eclipse v.13.6 for 6X
and 6XFFF beams using TRS483 CoP.
Relative output factors were also measured for
MLC collimation system to be compared with
that of the jaw collimation system. The
discrepancy of output factor between jaw-
shaped fields and MLC-shaped fields suggests
that jaw-based beam data itself may not
suitable for MLC-based treatment planning.
Additional measurement of small beam
percentage depth dose and profiles as well as
specific modelling of photon beam for MLC
system may be required.
REFERENCES
[1]. P. Andreo, “The physics of small megavoltage
photon beam dosimetry,” Radiother. Oncol.,
vol. 126, no. 2, pp. 205–213, 2018.
[2]. International Atomic Energy Agency, TRS483,
“An International Code of Practice for
Reference and Relative Dose Determination”,
November, 2017.
0.5
0.55
0.6
0.65
0.7
0.75
0.8
0.85
0.9
0.95
1
0 2 4 6 8 10
R
O
F
Field Size (cm, square equivalent)
6X TRS483 Output Factors
MLC
Jaws
Cone
0.5
0.55
0.6
0.65
0.7
0.75
0.8
0.85
0.9
0.95
1
0 2 4 6 8 10
R
O
F
Field Size (cm, square equivalent)
6XFFF TRS483 Output Factors
MLC
Jaws
Cone
DO DUC CHI et al.
55
[3]. I. J. Das, G. X. Ding, and A. Ahnesjö, “Small
fields: Nonequilibrium radiation dosimetry,”
Med. Phys., vol. 35, no. 1, pp. 206–215, 2008.
[4]. I. J. Das, J. Morales, and P. Francescon, “Small
field dosimetry: What have we learnt?,” in AIP
Conference Proceedings, 2016.
[5]. K. Park, J. Bak, S. Park, W. Choi, and S. W.
Park, “Determination of small-field correction
factors for cylindrical ionization chambers
using a semiempirical method,” Phys. Med.
Biol., vol. 61, no. 3, 2016.
[6]. H. Bouchard, J. Seuntjens, S. Duane, Y. Kamio,
and H. Palmans, “Detector dose response in
megavoltage small photon beams. I. Theoretical
concepts,” Med. Phys., vol. 42, no. 10, 2015.
[7]. E. Mainegra-Hing and B. R. Muir, “On the
impact of ICRU report 90 recommendations on
kQ factors for high-energy photon beams,” Med.
Phys., vol. 45, no. 8, pp. 3904–3908, 2018.
[8]. H. Benmakhlouf, J. Sempau, and P. Andreo,
“Output correction factors for nine small field
detectors in 6 MV radiation therapy photon
beams: A PENELOPE monte carlo study,”
Med. Phys., vol. 41, no. 4, 2014.
[9]. I. J. Das, M. B. Downes, A. Kassaee, and Z.
Tochner, “Choice of Radiation Detector in
Dosimetry of Stereotactic Radiosurgery-
Radiotherapy,” J. Radiosurgery, vol. 3, no. 4,
pp. 177–186, 2000.
[10]. M. Constantin, H. Hsu, S. Mansfield, and M.
Svatos, “Mitigation Strategies to Reduce the
Relative Extrafocal Dose ( RED ),” vol. 33, p.
2525, 2006.
[11]. K. Chełmiński and W. Bulski, “A multi-centre
analytical study of small field output factor
calculations in radiotherapy,” Phys. Imaging
Radiat. Oncol., vol. 6, no. September 2017, pp.
1–4, 2018.
[12]. W. Lechner et al., “A multinational audit of
small field output factors calculated by
treatment planning systems used in
radiotherapy,” Phys. Imaging Radiat.
Oncol., vol. 5, no. September 2017, pp. 58–
63, 2018.
[13]. Aspradakis, M. M. (2009). Small Field MV
Photon Dosimetry. World Congress on
Medical Physics and Biomedical Engineering,
September 7 - 12, 2009, Munich, Germany,
854–854.
[14]. Absorbed Dose Determination in External
Beam Radiotherapy: An International Code of
Practice for Dosimetry based on Standards of
Absorbed Dose to Water. Technical Reports
Series No. 398, International Atomic Energy
Agency, Vienna, 2000.
Các file đính kèm theo tài liệu này:
relative_output_factors_of_different_collimation_systems_in.pdf